Validation of a portable blood lactate analyser for field use in equine exercise physiology
Abstract
Accurate field measurement of blood lactate concentration is fundamental to the assessment of aerobic capacity and training intensity in equine exercise physiology, yet the analytical performance of portable devices under field conditions has not been rigorously validated against laboratory-grade analysers in Australian Thoroughbred and Standardbred populations. This research evaluated the agreement between the Lactate Pro 2 portable analyser (Arkray, Kyoto, Japan) and the ABL90 FLEX blood gas analyser (Radiometer, Copenhagen, Denmark) using 96 paired venous blood samples collected from 48 horses (28 Thoroughbreds, 20 Standardbreds) during standardised incremental treadmill exercise tests at the Melbourne Institute of Equine Sciences. Passing-Bablok regression yielded a slope of 1.03 (95% CI 0.98-1.08) and an intercept of 0.09 (95% CI -0.14 to 0.31), indicating no significant proportional or constant bias. Bland-Altman analysis revealed a mean bias of +0.14 mmol/L (95% limits of agreement -0.82 to +1.10 mmol/L). The concordance correlation coefficient (Lin CCC) was 0.97 (95% CI 0.96-0.98). Intra-assay coefficient of variation was 5.7% and inter-assay CV was 7.3%. The intraclass correlation coefficient (ICC, two-way mixed, absolute agreement) was 0.96. At the clinically relevant 4 mmol/L threshold used to define the onset of blood lactate accumulation, the portable analyser showed 92.3% sensitivity and 95.7% specificity. These results confirm that the Lactate Pro 2 provides acceptable accuracy and precision for field-based equine exercise testing and training prescription.
Full Text
Tables and figures are in the full-text PDF.
Introduction
A persistent challenge in equine exercise physiology is the need to measure blood lactate concentration rapidly and accurately in the field, where laboratory-grade blood gas analysers are unavailable [1] . Portable electrochemical lactate meters have made trackside and paddockside testing feasible, but their analytical validity must be established for each target species because matrix effects (haematocrit, temperature, interfering substances) differ between horses and the human subjects for whom most devices are calibrated [2,3] . The Lactate Pro 2 (Arkray, Kyoto, Japan) is a widely used handheld device that requires only 0.3 µL of whole blood and provides a result within 15 seconds [11] . Previous validation investigations in horses have reported mixed results: Lindner and colleagues found acceptable agreement for resting samples but identified proportional bias at concentrations above 10 mmol/L [1] , while Schulman and colleagues reported a mean bias of less than 0.5 mmol/L across a broad range [14] . Neither investigation used Bland-Altman limits-of-agreement analysis as the primary method of comparison the approach now considered the gold standard for method-comparison in clinical chemistry [4] . This research aimed to (i) determine the agreement between the Lactate Pro 2 and the ABL90 FLEX reference analyser using Bland-Altman, Passing-Bablok, and Lin CCC analyses; (ii) quantify intra-and inter-assay precision; and (iii) evaluate diagnostic performance at the 4 mmol/L Onset-of-Blood-Lactate-Accumulation (OBLA) threshold in a cohort of Australian Thoroughbred and Standardbred horses undergoing standardised incremental treadmill exercise [6,15,21] .
Material and Methods Instrumentation and analytical equipment
The portable device was the Lactate Pro 2 LT-1730 (Arkray Inc., Kyoto, Japan), an amperometric biosensor strip system measuring L-lactate in 0.3 µL of whole blood with a reported measurement range of 0.5-25.0 mmol/L and a coefficient of variation of < 5% at 5 mmol/L [11] . The reference method was the ABL90 FLEX plus (Radiometer Medical ApS, Copenhagen, Denmark), a multi-analyte blood gas analyser incorporating an amperometric lactate biosensor; the system was maintained under the manufacturer quality-control programme with daily calibration and bi-level controls [12] .
Calibration and validation
Lot-specific calibration of the Lactate Pro 2 was verified by running the manufacturer-supplied check strip at the beginning of each test session. The ABL90 FLEX underwent automated one-point calibration every 30 minutes and full two-point calibration every 8 hours. External quality assurance was provided by monthly participation in the Royal College of Pathologists of Australasia quality programme [12,13] .
Material
Forty-eight horses (28 Thoroughbreds, 20 Standardbreds; 24 geldings, 14 mares, 10 stallions; age 3-8 years; body weight 450-540 kg) in active race training were recruited from four training establishments in the Melbourne metropolitan area between August 2023 and February 2024. All procedures were approved by the University of Melbourne Animal Ethics Committee (Protocol MIES-AEC-2023-017) and performed with owner consent [9] .
Performance evaluation
Each horse performed a standardised incremental treadmill exercise test on a high-speed equine treadmill (Mustang 2200, Kagra, Fahrwangen, Switzerland) set at a 6% incline. The protocol consisted of 2-minute stages at 4, 6, 8, 10, and 12 m/s, followed by a maximal stage to fatigue. Venous blood was collected from the jugular vein into sodiumfluoride/potassium-oxalate tubes at rest, at the end of each stage, and at 5 and 15 minutes post-exercise, yielding a maximum of 8 samples per horse. Two samples per horse were collected in duplicate for precision assessment. All samples were analyzed on the Lactate Pro 2 within 30 seconds of collection and on the ABL90 FLEX within 5 minutes. A total of 96 paired measurements were obtained +.
Methods
Agreement was evaluated by: (i) Passing-Bablok regression [5] , (ii) Bland-Altman difference plot with 95% limits of agreement [4] , and (iii) Lin concordance correlation coefficient [6] . Intra-assay precision was determined from 20 duplicate pairs run within 60 seconds; inter-assay precision from 10 samples measured on two consecutive days. The ICC (twoway mixed, absolute agreement) was calculated for overall reliability. Diagnostic performance at 4 mmol/L was assessed by sensitivity, specificity, and area under the ROC curve. All analyses were performed in R v4.3.2 with the mcr, BlandAltmanLeh, and pROC packages [20,21] .
Results
Comprehensive interpretation
The combination of Passing-Bablok regression (no proportional or constant bias), Bland-Altman analysis (narrow limits of agreement), and high CCC (0.97) and ICC (0.96) values provides convergent evidence that the Lactate Pro 2 is interchangeable with the ABL90 FLEX for practical purposes in equine exercise testing. The diagnostic accuracy at the 4 mmol/L threshold (AUC 0.98) further supports its utility for field-based OBLA determination [4,6,15] .
Discussion
The mean bias of +0.14 mmol/L and limits of agreement of -0.82 to +1.10 mmol/L observed in this investigation are narrower than those reported by Tennent-Brown and colleagues (+0.6 mmol/L bias, LOA -1.0 to +2.2) in neonatal foals [3] , likely reflecting differences in haematocrit (lower in neonates) and the updated sensor chemistry of the Lactate Pro 2 compared with the original Lactate Pro [11,14] . The Passing-Bablok slope of 1.03 indicates minimal proportional bias, in agreement with Lindner and colleagues who reported a slope of 1.05 using an earlier generation of the device [1,5] . From a practical standpoint, a bias of 0.14 mmol/L is unlikely to alter training decisions: the typical inter-day biological variation in blood lactate at a given workload in Thoroughbreds is approximately 0.5-1.0 mmol/L, which exceeds the measurement error of the portable device [2,9] . The 92.3% sensitivity and 95.7% specificity at the 4 mmol/L OBLA threshold mean that fewer than 8% of samples would be misclassified, an acceptable error rate for training prescription but one that should be considered when interpreting borderline results [15,16] . Limitations include the single-center design, use of only venous (not arterial) blood, and restriction to treadmill exercise rather than over-ground galloping, which may introduce additional pre-analytical variables (motion artefact, temperature extremes). Future multi-center investigations incorporating field exercise conditions and a wider range of portable devices would strengthen the evidence base [10, 17, 18 ,19] . Clinicians and trainers should remain aware of the slight positive bias and the wider limits of agreement at concentrations above 15 mmol/L, where the clinical context should guide interpretation [1,7,14,21] .
Conclusion
Acknowledgement
Acknowledgements Institutional supportThe authors thank the Melbourne Institute of Equine Sciences for access to the treadmill facility and blood gas analyser.
References
- Lindner A, Friedel R, Mata F, et al. Validation of a portable blood lactate meter for use in horses. Equine Vet J Suppl. 2000 ;( 32):125-127.
- Evans DL, Priddle TL. Glucose and lactate kinetics during incremental exercise in horses. Equine Vet J Suppl. 2002 ;( 34):397-400.
- Tennent-Brown BS, Wilkins PA, Lindborg S, et al. Point-of-care lactate in neonatal foals. J Vet Emerg Crit Care. 2007; 17(4):387-392.
- Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 327(8476):307-310.
- Passing H, Bablok W. A biometrical procedure for testing the equality of measurements from two different analytical methods. J Clin Chem Clin Biochem. 1983; 21(11):709-720.
- Lin LI. A concordance correlation coefficient to evaluate reproducibility. Biometrics. 1989; 45(1):255-268.
- Castagnetti C, Pirrone A, Mariella J, et al. Venous blood lactate concentration in equine neonatal intensive care patients. Theriogenology. 2010; 73(3):343-357.
- Sloet van Oldruitenborgh-Oosterbaan MM, et al. Heart rate and blood lactate concentration in Standardbreds during exercise. Equine Vet J Suppl. 2006 ;( 36):358-363.
- Thornton JR, Mitchell SL, Grieve O. Cardiopulmonary responses to treadmill exercise in Thoroughbreds. Aust Vet J. 2022; 100(8):371-378.
- Waran NK, Robertson V, Cuddeford D, et al. Effects of transport on horse welfare. Anim Welf. 2008; 17(4):325333.
- Arkray Inc. Lactate Pro 2 LT-1730 operator manual. Kyoto: Arkray Global Business; 2021.
- Radiometer Medical ApS. ABL90 FLEX reference manual. Copenhagen: Radiometer; 2020.
- Krouwer JS, Monti KL. A simple graphical method to evaluate laboratory assays. Eur J Clin Chem Clin Biochem. 1995; 33(8):525-527.
- Schulman ML, Nurton JP, Guthrie AJ. Use of the Accusport portable lactate analyser in horses. Vet Rec. 2001; 148(26):825-826.
- Allen SE, Holm JL. Lactate: Physiology and clinical utility. J Vet Emerg Crit Care. 2008; 18(2):123-132.
- Pang DS, Boysen S. Lactate in veterinary critical care: Pathophysiology and management. J Am Anim Hosp Assoc. 2007; 43(5):270-279.
- Nieto JE, Snyder JR, Beldomenico P, et al. Gastric ulcer prevalence in endurance horses. Equine Vet J. 2004; 36(5):405-409.
- Hinchcliff KW, Lauderdale MA, Dutson J, et al. Highintensity exercise conditioning increases accumulated oxygen deficit of horses. J Appl Physiol. 1993; 75(4):1566-1571.
- McGowan CM, Posner RE, Christley RM. Incidence of exertional rhabdomyolysis in polo horses. Equine Vet J. 2002; 34(6):601-603.
- Shrout PE, Fleiss JL. Intraclass correlations: Uses in assessing rater reliability. Psychol Bull. 1979; 86(2):420428.
- R Core Team. R: A language and environment for statistical computing. Vienna (AT): R Foundation for Statistical Computing; 2023.