A framework for evaluation of deformable image registration spatial accuracy using large landmark point sets

Physics in Medicine and Biology · 2009 · Vol. 54(7) · pp. 1849–1870
Richard CastilloEdward CastilloRudy GuerraValen E. JohnsonTravis McPhailAmit GargThomas Guerrero

Abstract

Expert landmark correspondences are widely reported for evaluating deformable image registration (DIR) spatial accuracy. In this report, we present a framework for objective evaluation of DIR spatial accuracy using large sets of expert-determined landmark point pairs. Large samples (>1100) of pulmonary landmark point pairs were manually generated for five cases. Estimates of inter- and intra-observer variation were determined from repeated registration. Comparative evaluation of DIR spatial accuracy was performed for two algorithms, a gradient-based optical flow algorithm and a landmark-based moving least-squares algorithm. The uncertainty of spatial error estimates was found to be inversely proportional to the square root of the number of landmark point pairs and directly proportional to the standard deviation of the spatial errors. Using the statistical properties of this data, we performed sample size calculations to estimate the average spatial accuracy of each algorithm with 95% confidence intervals within a 0.5 mm range. For the optical flow and moving least-squares algorithms, the required sample sizes were 1050 and 36, respectively. Comparative evaluation based on fewer than the required validation landmarks results in misrepresentation of the relative spatial accuracy. This study demonstrates that landmark pairs can be used to assess DIR spatial accuracy within a narrow uncertainty range.

Medical Image Segmentation TechniquesAdvanced Vision and ImagingMedical Imaging and AnalysisLandmarkComputer scienceRange (aeronautics)Standard deviationArtificial intelligenceImage registrationGround truthPoint (geometry)Accuracy and precisionSample (material)

MeSH terms

AlgorithmsEsophageal NeoplasmsHumansImage Processing, Computer-AssistedReference StandardsSensitivity and SpecificityRadiography, ThoracicTomography, X-Ray ComputedReproducibility of ResultsUncertainty

Funding

  • National Institutes of Health
  • National Cancer Institute
Citations
584
FWCI
23.92
field-weighted impact
References
38
Percentile
100%
vs. same field & year
Citations per year
Cited by
References
Image deformation using moving least squares
ACM Transactions on Graphics · 2006 · 676 citations
The computation of optical flow
ACM Computing Surveys · 1995 · 1,251 citations
Medical image registration
Physics in Medicine and Biology · 2001 · 2,073 citations
Numerical recipes in C: the art of scientific computing
Choice Reviews Online · 1993 · 17,990 citations
<title>Determining Optical Flow</title>
Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE · 1981 · 7,482 citations
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.