A Biomechanical Study of Intrapeduncular Screw Fixation in the Lumbosacral Spine

Clinical Orthopaedics and Related Research · 1986 · Vol. 203(203) · pp. 99–112
Michael R. ZindrickLeon L. WiltseERIC H. WIDELLJames C. ThomasW. RUSSELL HOLLANDB. TED FIELDCurtis W. Spencer

Abstract

This laboratory experiment was undertaken to identify factors contributing to intrapeduncular screw fixation in the vertebra. Testing was performed in axial pull-out and cyclic loading modes using multiple screw designs inserted to various depths into fresh human lumbosacral vertebra. The degree of osteoporosis played a major role in pull-out strength. Larger diameter, full-threaded screws inserted deep enough to engage the anterior vertebral cortex resulted in the most secure fixation. In the sacrum, the second sacral pedicle was the weakest location of insertion. Screws aimed laterally into the ala at 45 degrees or medially into the first sacral pedicle resisted larger axial pull-out loads than those inserted straight anteriorly into the ala. Methyl methacrylate was found to restore secure fixation in previously-loosened screws and pressurization of cement doubled the pull-out force. In cyclic load tests, deeper-inserted screws were found to withstand a greater number of cycles before loosening. Measurements of pedicle outer cortical diameters were found in many specimens to be smaller than both the 4.5-mm and 6.5-mm diameter screws.

Spinal Fractures and Fixation TechniquesPelvic and Acetabular InjuriesSpine and Intervertebral Disc PathologySacrumMedicineLumbosacral jointFixation (population genetics)VertebraAnatomyBiomechanicsOrthodonticsPopulation

MeSH terms

AgedBiomechanical PhenomenaBone ScrewsHumansLumbar VertebraeMiddle AgedSacrumSpinal Fusion
Citations
641
FWCI
2.11
field-weighted impact
References
0
Percentile
87%
vs. same field & year
Citations per year
Citation Network

How this paper connects to the literature. Drag to explore, click any node to open that paper.